Image Quality & Information Technology


Image Quality & Information Technology

medphy



Learning Objectives

  • Understand what “good clinical image quality” means
  • Learn about the three key technical IQ metrics (spatial resolution, contrast and noise), how these can be measured and the impact on Image Quality (IQ)
  • Understand advantages and limitations of digital images
  • Identify the contents of DICOM clinical image files and the typical size/archiving requirements
  • Learn about the contrast limitations of the human visual system and the value of post-processing tools.
  • Calculate the size of image files from given detector specs, data and image characteristics
  • Apply the concept of Nyquist frequency to digital imaging problems.
  • Overview of Information technologies (IT) in healthcare, including Picture Archive and Communication Systems (PACS) and Electronic Health Records (EHR).

Medical Imaging Modalities

There are many different imaging modalities:

  • Ultrasound (US): Sound waves (mechanical energy), are very good at showing fluid (a dark echo will be seen).
  • Computed Tomography (CT): Uses X-Rays at higher energies. Primarily provides information about the anatomy.
  • Positron Emission Tomography - Computed Tomography (PET-CT): Positrons and gamma photons from PET, and X-rays from CT.
  • Magnetic Resonance Imaging (MRI): uses magnetic fields and radio-frequency. Primarily provides information about the physiological.
  • Positron Emission Tomography (PET): uses positrons and gamma photons. Primarily provides information about the physiological.
  • Fluoroscopy: uses X-rays
  • Mammography: uses X-rays at lower energies as breast is fat, soft tissue. Reduce energy to amplify photoelectric effect vs Compton scatter. Tomosynthesis is an advanced application of mammography to see images with less overlap of structure of the breast. Primarily provides information about the anatomy.
  • Positron Emission Tomography - Magnetic Resonance (PET-MR): uses positrons and gamma photons from PET; magnetic fields and radio-frequency from MRI.

Image Quality IQ

Image Quality (IQ) is a general and subjective concept best described within the context of the specific task. An image with a good IQ has suitable characteristics for the intended use which could be screening, diagnostic, intervention or follow up.

Note: IQ does not mean aesthetically beautiful images!

For example in breast imaging, a high image quality enables detection and characterisation of:

  • Micro-calcifications in clusters
  • Nodules which are more dense than surrounding tissue
  • Architectural distortions helps radiographers to how symmetric breasts are to each other
  • Cysts, fluid
  • Angiogenesis, the blood supply (shown by MRI)
  • Increased glucose metabolism, important as often associated with cancer and other pathology (shown by PET)

Ideally with a high (100%) sensitivity (the ability to correctly identify the structures, the true positive rate) and a high (100%) specificity (the ability to correctly identify the structures without disease, the true negative rate).

Sensitivity Equation

$ Sensitivity = \frac{True \ Positive}{True \ Positive \ + \ False \ Negative} \times 100 $

Specificity Equation

$ Specificity = \frac{True \ Negative}{True \ Negative \ + \ False \ Positive} \times 100 $

Factors that affect IQ

Image quality is affected by information content, perception/interpretation and decisions by the observer:

Information content
  • Tissue characteristics and pathology
  • Radiographic technique (e.g. positioning, compression)
  • Equipment specification (e.g. pixel size, dynamic range)
  • Equipment performance (e.g. AEC setup, noose characteristics, etc)
  • Post-processing (noise reduction, edge-enhancement)
Perception & Interpretation
  • Viewing conditions (e.g. ambient light)
  • Monitor specs (matrix size, pixel size, bit depth)
  • Visual acuity
Observer decision criteria
  • A priori knowledge
  • Experience
  • Personal Preference

Routine quality control aims to monitor equipment performance over time and compare it with a baseline/reference to ensure it adheres to the intended standards through the lifetime of the equipment.

ALARA / ALARP As Low As Reasonably Achievable/Practicable

Modalities involving ionising radiation require image quality to be complaint with ALARA/ALARP.

Digital Imaging Systems and Patient Dose

Analogue systems using film were sensitive to an upper and a low threshold. Too low a dose would result in the film being underexposed and too high a dose would result in the film being over exposed.

However digital systems (CR and DR) have wider dynamic range and are tolerant to sub-optimal exposure conditions. Therefore it is very hard for the operator to distinguish whether the machine is malfunctioning and potentially the patient could receive too little or too much radiation dose (dangerous!).

Digital Images

A digital image is an array of numbers assigned to each pixel or voxel. In a digital image the picture is broken down into discrete blocks. In a 2D system each block is termed a pixel (picture element) and in a 3D system each block is termed a voxel (volume element). A digital image is numerically described by:

Array Size 4 x 4 = 16 pixels

The array size determines the sampling frequency (pixels/mm). The higher the sampling frequency the better the representation of the object detail.

Bit Depth 2 bit = 22 = 4 possible pixel values

The bit depth determines the number of possible values that can be assigned to a pixel. Quoted as the number of bits allocated to the image, so the simplest image would be 1-bit = 21 = 2 possible values = black & white.

Terminology
1 bit 1 binary digit
2 nibble 4 bits
1 byte 8 bits
1 word 2 bytes (generally)
1 kilobyte 210 = 1024 bytes
1 megabyte 220 = 1024 kbytes
1 gigabyte 230 = 1024 Mbytes

The Human Visual System (HVS) is a little under 8-bit i.e. can distinguish ~200 Just Noticeable Differences (JND) in grey scale level. Medical imaging detectors and displays are typically 12-bit (i.e. 4096 grey levels) as post-processing tools manipulate and optimise the image for HVS.

Note: The representation of an object improves as the array size and bit depth are increased.

What is the image representation if the size is 2300 x 1900 at a 16 bit depth?

Image Size = 2300 x 1900 x 2 byte per pixel
Image Size = 8 740 000 bytes
Answer = 8.3 Mb

Digital image presentation

Digital images are normally viewed:

  • Hardcopy (film) using a light box
  • Medical Display (CRT or LCD) usually 3 MegaPixels upwards (5.9 MegaPixels for mammography)

Image processing tools

The same image at different window width and level settings show different information. Post processing may generate artefacts in the image (e.g. high level of edge enhancement may suggest that an implant is loose).

Compression and image data

Lossy compression can make a file a lot smaller, however it is required by law that medical images have a lossless compression (to avoid any degradation in quality that could cause a change in diagnosis).

Advantages of Digital Images Disadvantages of Digital Images
Wide Dynamic range Lower spatial resolution (Still may be adequate for clinical task)
Post processing capabilities Initial cost can be high
Portability & telemedicine Users have to monitor dose/patient exposure closely
Security & backup  
Less physical storage space required  
Advanced applications (CAD, Image subtraction, tomosynthesis, etc)  
Clean and safe processing  

DICOM format

Medical images are usually in the Digital Imaging and Communications in Medicine (DICOM) format. A DICOM file has two components:

  1. Clinical information (signals with a clinical meaning)
  2. Acquisition and image info (in the DICOM header)

All electronic detectors produce an analogue signal which varies continuously and which depends on the amount of radiation (or other form of energy) received by the detector. In most modern electronic imaging systems, the analogue system from the detector is transformed into a digital signal, that is a signal that has a discrete, rather than continuous values. During this transformation obviously some information is lost.

TYPICAL SIZE OF MEDICAL IMAGES
Study Archive capacity required (uncompressed Mb)
Chest X-ray (PA + L, 2 x 2 kby) 20
CT series (120 images, 512 x 512) 64
SPECT myocardial perfusion study (TI 201) 1
US study (60 images, 512 x 512) 16
Cardiac catheterisation 450 - 3000
Mammogram (screening) 2x CC + 2x MLO 32 - 220

Technical descriptors of IQ

Spatial resolution, contrast and noise are the three key indicators of Image Quality. From these descriptors, Signal-to-Noise Ratio (SNR) and Contrast-to-Noise Ratio (CNR) can be derived. When measured under controlled conditions these can be very useful values.

Signal-to-Noise Ratio SNR

SNR shows how many times stronger the signal is compared to the noise (signal variations). If all the sources of non-random noise can be removed than then the dominant source of noise is random (Poisson) distribution.

$ SNR = \frac{signal}{noise} = \frac{signal}{\sqrt{signal}} = \sqrt{signal} $

Contrast-to-Noise Ratio CNR

CNR is a useful metric in medical imaging as it allows us to quantify subtle variations in signal between objects and their surrounding background.

$ CNR = \frac{ \vert signal_{obj} \ - \ signal_{bkgd} \vert }{noise_{bkgd}} $

Its best to have a high photoelectric absorption and low Compton scatter. Important requirements of an imaging system is that is has a high signal detection efficiency with a high SNR and a high CNR.

1. Spatial resolution

An ideal detector would produce an exact representation (sharp response) of the object irrespective of the spatial frequency. However in reality the response is more curved. Spatial resolution affects the visibility of detail in an image and the ability to detect small structures close to each other. Poor spatial resolution of the imaging system shows a blur in the image. Decreasing the pixel size improves the spatial resolution at the cost of more noise as there are less photons per area (unless you increase the dose to compensate for this).

Sampling frequency

Nyquist-Shannon’s Sampling Theorem states if you have a signal that is perfectly band limited to a bandwidth of f0 (cycles/mm) then you can collect all the information there is in that signal by sampling it at discrete times, as long as your sample rate is greater than 2f0 (samples/mm)

For example, if the maximum frequency in the object is 2cycles/mm then the sampling must be done at least 4 cycles/mm.

$ N_F = \frac{1}{2 \times Pixel \ Pitch} $

Where the pixel pitch is the distance between two adjacent pixels.

Under-sampling occurs at a sample rate below the Nyquist rate. This leads to misrepresentation of the signal, loss of information and generation of artefacts.

Imaging Modality Pixel Size Nyquist frequency lp/mmm
Mammography > 0.080 m 6.3
General Radiography > 0.143mm 3.5
Fluoroscopy > 0.200mm 2.5

2. Contrast Z dependent

Contrast key to detect subtle signals and is determined by the relationship between the magnitude of the signal and the magnitude of the fluctuations in the signal (noise). It depends on the composition and thickness of an object as well as the properties of the detector such as noise.

3. Noise

An ideal imaging system would:

  • Detect all X-rays
  • Preserve all spatial information
  • Absorb all energy from each x-ray
  • No additional noise present in the system

However no such detector exists and noise is fashioned which prevents the visibility of small/low contrast details. Some sources of noise include:

  • Gain calibration is where digital detectors can compare a raw image to reference values to compensate and produce a uniform image. This can produce electronic noise.
  • Mottle (quantum noise) is determined by the stochastic (random) nature of X-ray production (a process we can’t control). Ideally X-ray systems operate in a quantum limited regime. i.e. where quantum noise is the limiting noise source and quantum noise decreases with increasing number of photons.
  • Clutter (anatomic noise) is where the anatomy of the human body disrupts the region to be viewed. For example in chest radiography the detection of subtle lung nodules is limited by anatomic noise. Another example is the superimposition of breast tissue in mammography which degrades IQ and poses difficulty to lesion detection.
  • Other sources include, electronics, detector structure/defects and quantisation (restricting the number of values of a system)

Noise can be reduced, but never eliminated completely. CNR provides valuable data to investigate drops in Image Quality.

Health Information technology HIT

HIT has changed the way healthcare is provided. It holds great promise towards improving healthcare quality, safety and costs. Some examples of IT in healthcare:

  • Picture Archiving and Communication Systems (PACS)
  • Electronic Health Record (EHR)
  • Electronic prescription services
  • Hospital Information Systems (HIS)
  • Radiology Information Systems (RIS)
  • Incident alert systems
  • Patient registers
  • NHSmail

Picture Archiving and Communication Systems PACS

There were some key milestones in the development of PACS:

  • 1979: First digital data link between CT scanner and radiation treatment planning computer (Loma Linda University Medical Centre, California)
  • 1993: PACS implemented in Hammersmith Hospital, UK
  • 1993: DICOM 3.0 (originally ACR-NEMA 3.0) standard published
  • 1996: The first filminess hospital in operation in the UK (Hammersmith Hospital)
  • 1998: Integrating the Healthcare Enterprise (IHE) initiative established. …

PACS continue to develop, with technological advances making implementation similar and cheaper. Much current development focus on workflow and systems integration. At the moment, PACS typically comprises of:

  • Data storage devices
  • Image display devices
  • Software
  • Film printers and digitisers
  • Computer networks

May have additional networks to the other IT systems (HIS, PAS, RIS)

Benefits of PACS
  • Less physical space required
  • Easy image access
  • Safety
  • Efficient data management
  • Cost savings
  • Environmental benefits
  • Enables teleradiology
Challenges of PACS
  • High capital investment and ongoing costs
  • Integration with other (local and remote) IT
  • Continuous user training
  • Quality assurance
  • Cost
  • Specialised management/Technical skills

Electronic Health Record (EHR)

This is a record of important clinical information about the patient, and provides key performance indicators for the hospital or specialist unit (e.g. to support research, help planning new services):

  • Consultation notes
  • Hospital admission records and reasons
  • Chronic health conditions (e.g. diabetes, asthma)
  • Test results (x-ray, CT, MRI) and images
  • Radiation dose received in imaging procedures
  • Treatment received, medicines taken
  • Adverse reactions to medications
  • Hospital discharge records, follow-up appointments
  • Lifestyle information (e.g. smoking)
  • Personal details (NHS number, age, gender, address)

The EHR can be created, managed an consulted by authorised providers and staff across more than one health care organisation. It can bring together information from current and past doctors, emergency facilities, school and workplace clinics, pharmacies, laboratories and medical imaging facilities.

The UK shows the biggest take-up of electronic health records in Europe. $2.1 billion (4% annual growth) was spent by the UK by the end of 2015 compared to $9.3 billion (7.1% annual growth) spent by the US.

Impact of EHR

The top 10 functions where doctors globally perceive a positive impact of EMR and HIE:

  • Improved co-ordination of care across care settings/service boundaries
  • Improved health outcomes
  • Increased speed of access to health services
  • Reduced number of unnecessary interventions/procedures
  • Improved patient access to specialist health care services
  • Reduction in medical errors
  • Better access to quality data for clinical research
  • Improved cross-organisational working processes
  • Improved quality of treatment decisions
  • Improved diagnostic decisions

However there are some challenges of implementing EHR. Potential of EHRs meets problems of implementation as they could distract from doctor-patient relationships, wasting valuable time and driving up costs (costly to maintain).

QUESTIONS


What does “good IQ” means in the context of medical images?
A good IQ has suitable characteristics for the intended use which could be screening, diagnostic, intervention or follow up.

What factors that influence IQ?…and perceived IQ?
Answer coming soon

What are the 3 key technical descriptors of IQ?
Answer coming soon

What are the main sources of noise in X-ray imaging? And their causes?
Answer coming soon

How can CNR be measured? What affects it?
Answer coming soon

How can the performance of a medical monitor be assessed?
Answer coming soon

What data is contained in a DICOM file?
Answer coming soon

How does image matrix size and bit depth affect image quality?
Answer coming soon

What differences are expected between an 12bit and a 8bit image?
Answer coming soon

How does SNR relates with the number of photons used to produce an X- ray image for an ideal x-ray imaging system?
Answer coming soon

What is spatial resolution and how can it be improved for a digital system?
Answer coming soon

Discuss advantages and limitations of digital imaging systems?
Answer coming soon

What are the 2 main functions of an Electronic Health Record (EHR)?
Answer coming soon

Give examples of impact of EHR on patient and the healthcare system.
Answer coming soon

What is PACS?
Answer coming soon

How can PACS affect workflow in the imaging department?
Answer coming soon

Discuss key requirements of a hospital PACS?
Answer coming soon

How could IT systems support the management of adverse incidents in a hospital setting?
Answer coming soon

Discuss the introduction of IT technologies in healthcare and how they can bring benefits to patients and the healthcare system?
Answer coming soon

PROBLEMS


In the plane of the detector what spatial frequency can be recorded by a 512 x 512 pixel digital fluoroscopy system with 150mm x 150mm receptors?

Detector size = 150mm x 150mm
Matrix size = 512 pxls x 512pxls
Pixel pitch (d) = 150mm/512 = 0.293 mm
Nyquist Frequency (Nf) = 1/2.d = 1 /2(0.293) = 1.71 l p/mm
Solution: 1.71 lp/mm


A grayscale chest radiograph is 35cm x 29cm in area and was digitised with a sampling frequency that preserves the inherent spatial resolution in the image which is approximately 5 lp/mm (line pairs per millimetre). Each sample was digitised with 16 bits.

(a) Determine the image array size (in pixels)
The minimum pixel size to preserve the frequency is calculated using the Nyquist theorem:
Nf=1/2p … p=1/(2x5 lp/mm) = 0.1 mm
Image array size (pixels) = 350/0.1 x 290/0.1 = 3500 pxl x 2900 pxl

(b) Calculate the memory (in Megabytes) required to store a chest radiograph composed of an antero-posterior (AP) and a lateral (L) view of the chest (i.e. 2 images)
Memory required for one image
= 3500 x 2900 x 2 = 20 300 000 bytes / 1024 bytes/kbytes
= 19 824 kbytes / 1024 kbytes/Mbytes
= 19.3 Mbytes
Memory required for AP + Lateral ~39 Mbytes


return  link
Written by Tobias Whetton